Temperature uniformity test of freeze dryer plate layer

(1) Pre-calibration. Before the verification, put the verification temperature probe and the standard temperature probe into the temperature dry well at the same time, and perform pre-calibration. , the calibration reading deviation should be < 0.5 ℃.

(2) Connect the calibrated temperature probe into the freeze dryer through the verification port, place the 1-23# temperature probe, the numbers 1-5 represent the freeze-dryer product layer, and T1-3# is the temperature probe placed on the first The silicone oil inlet, outlet and center position of 3 layers, and other temperature probes are placed at the 4 corners and center positions of each layer. Start the freeze dryer, set the temperature of the heat transfer oil to 3 points of 40°C, 0°C and 40°C, respectively. After the inlet and outlet temperatures of the heat transfer oil reach equilibrium at each set temperature point, run for 30 min, and keep at -40 The uniformity of the layer temperature at ℃, 0℃ and 40℃. Perform 3 replicate tests. After the verification test is completed, the temperature probe will be used for post-calibration. The calibration points are set to 3 points of -40°C, 0°C and 40°C, and the reading deviation of the post-calibration should be <0.5°C.

(3) Eligibility criteria. According to the national pharmaceutical machinery industry standard JB T20032--2012 "Medical Vacuum Freeze Dryer", and combined with the product process requirements, when kept at 40 °C, 0 °C and 40 °C, all test points of each board layer have the highest temperature at the same time. The temperature difference between the value and the minimum value should be ≤2℃, and the uniformity of the board layer is qualified.


You Might Also Like

Send Inquiry