Freeze dryer online cleaning CIP coverage
(1) Spray a layer of vitamin B2 aqueous solution on the inner surface of the entire freeze-drying chamber with a concentration of 15 mg/L. Pay special attention to the parts that are difficult to clean (such as the nozzle, the top of the box and the bottom of the board) to be completely sprayed. Turn on the water for injection and start the CIP cycle. After completing the CIP, use a fluorescent lamp to illuminate the inner surface of the cavity to find out whether there is any residual vitamin B or fluorescent substance, and repeat the test 3 times.
(2) Eligibility criteria. After CIP cleaning, there is no visible fluorescent substance on the inner surface of the cavity, and the cleaning coverage rate is 100% (refer to the industry standard of pharmaceutical vacuum freeze dryer JB/T20032-2012).